Fully Automatic Current Transformer Testing System ZCHG-A6

ZCHG-A6 Fully Automatic Current Transformer Testing System is a new generation verification device developed by our company to meet the requirements for rapid and accurate modern transformer calibration. The system comprises several components: the ZC-103B Intelligent Transformer Calibrator, ZC-113B Current Load Box, ZC-506 Dual-stage Programmable Voltage Regulator, ZC-430 Insulation Resistance Tester, the ZCHG-A6 automatic verification unit, and a computer. Operators use the dedicated error verification management software to configure key parameters—such as transformer type, measurement range, accuracy class, secondary burden, power factor, and test positions. Utilizing RS485 communication and control, the system can test six current transformers with identical ratios in a single current-ramp cycle, significantly enhancing operational efficiency.

Features and Advantages

1. The system supports three testing modes: online automatic, standalone automatic, and standalone manual.

2. It utilizes an RS485 communication port, ensuring stable signal transmission and strong interference immunity.

3. Primary and secondary wiring connections for the standard current transformer are automatically switched based on computer-configured parameters.

4. Voltage and current load box settings are switched via programmable control, eliminating the need for manual calculations or gear changes.

5. Test positions can be selected arbitrarily, with the active position displayed in real-time during testing.

6. The high-speed transformer calibrator features comprehensive protection functions, including over-current and over-voltage protection, as well as detection for polarity errors and signal faults.

7. The error management software records, prints, and backs up test data in compliance with verification regulations and allows for the customization of certificate formats.

8. The error management software supports seamless data integration with the State Grid SG186 marketing system (relevant documentation required).

9. The system features an aesthetically pleasing design, user-friendly functionality, and automated status displays, creating an optimized working environment.

Product Picture

  • ZCHG-A6 Current Transformer Test System 1
  • ZCHG-A6 Current Transformer Test System 3
  • ZCHG-A6 Current Transformer Test System 2

Parameter

  • Environmental conditions for the use of CT testing system
  • CT Testing system parameters
  • Programmable CT Load boxTechnical Parameters
Environmental conditions for the use of CT testing system
Temperature-10℃~50℃
Relative humidity<80% (25℃)
Elevation<2500m
Power frequency50Hz/60Hz±10%
Power supply voltage (control loop and instrument power supply)220V±10%
CT Testing system parameters
Working rangeCurrentPrimary current5~2000A
Secondary current5A
Accuracy Class0.02(S)
Measuring rangeIn-phase component (%)0.0001 ~ 200.0
Resolution0.0001
Orthogonal component (min)0.001 ~ 700.0
Resolution0.001
Impedance (Ω)0.0001 ~ 20.0
Resolution0.0001
Admittance (ms)0.0001 ~ 20.0
Resolution0.0001
Programmable CT Load boxTechnical Parameters
Rated LoadCurrent2.5VA~60VA COSΦ=0.8(selectable)
2.5VA~10VA COSΦ=1.0(selectable)
Measuring range and accuracyCurrent1%~120%(optional 200%)
Secondary current5A
Accuracy3% ±0.05VA