Features and Advantages
1.Our instrument has designed in integrated function of oil cup, heating, temperature control and voltage regulating
2.Displayed with color large LCD screen, simple to operate
3.Empty cup self-calibration
4.Equip with over-voltage, over-current and temperature limitation protection
5.Medium frequency induction heating electrode cup and short time uniform heating
6.Direct measure temperature with probe placed inside the measuring electrode cup
7.Contains sine-wave-generator, high power digital voltage regulator test power supply can provide 50Hz standard voltage
Parameter
| Condition of Use | -5℃∽40℃ RH<80% | |
| Power Source | AC 110V±10% Without Frequency Limit | |
| AC High Voltage Output | 400V∽2200V ±2% Every 100V 50VA | |
| DC High Voltage Output | 200V∽600V ±2% | |
| Temperature Control Induction Stove | Maximal Power 500W | |
| Temperature Control Range | <110℃ | |
| Temperature Measuring Error | ±0.5℃ | |
| Control Temperature Error | 0.1℃ | |
| Control Temperature Time | Room Temperature 100℃ <20min | |
| Measuring Range | tgδ | Without Limit |
| Cx | 15PF-300PF | |
| Rx | 10M-20T | |
| Resolution | tgδ | 0.001% |
| Cx | 0.001pF | |
| Rx | 0.01 | |
| Precision | △tgδ | ±(Reading*0.5%+0.020%) |
| △Cx | ±(Reading*0.5%+0.5PF) | |
| △Rx | ±Reading * 10% | |
| Relative Dielectric Loss Constant | er | Calculated Automatically |
| Oil DC resistance | Rx | Calculated Automatically |
| Oil DC conduction | Gx | Calculated Automatically |
| Volume Resistivity | ρ | Calculated Automatically |
| Bulk conductivity | σ | Calculated Automatically |
| Dimension | 410(L)×310(W)×305(H) | |
| Weight | 18 Kg | |